Support About us Installations News Contact us Find us Glossary
Home
Our products
PROTOC ®
PROTOC ® TL
LABTOC ®
PROAM
TERMINATOR
TOC - our process
NH4-N-our process
Find by parameter
Find by application
Other products
Site map


PROTOC ®

Product Literature


PROTOC® is a fast, reliable and flexible on-line analyser system designed to accurately determine TOC/TC concentrations in real time. The analyser comprises a wall mount analyser section typically supplied complete with a controller, providing local display and outputting signals.

  • Protoc® 100
    Low cost analyser with manually operated valve to calibrate/zero/clean
  • Protoc® 300
    Economical automated analyser for single stream analysis (MCERTS model)
  • Protoc® microSpyder and Web(s)
    Economical multi-point 1-4 stream analysis with touch screen controller
  • Protoc® Spyder and Web(s)
    Advanced multi-point 1-4 stream analysis with touch screen controller

PRINCIPLE OF OPERATION

PROTOC® instruments use an Ultra-Violet promoted persulphate oxidation to continuously determine the contamination of organic chemicals present in the sample. A carrier gas continuously sparges the reaction vessel liberating the resultant CO2 gas which is delivered to an Infra-Red detection system. Inorganic carbonates can be automatically removed by sample pre-treatment using an acid sparge

 




CONTROL AND MONITORING OPTIONS
Various models are available offering different levels of automation. Two point automatic calibration (at zero and typically at 50% of the measuring scale), maintains accuracy and validates measurement performance. Provision of an acid wash can also enhance the self-cleaning measuring technique on the more challenging applications. Various options are also available for data display, data logging and telemetry communications. The Protoc® 300 model has been certified to meet the requirements of MCERTS (certificate SIRA MC 060077/00).


TYPICAL SPECIFICATIONS

Measuring Range:

0-5 up to 50,000 PPM (with dilution)

Repeatability:

+/- 2% or better

Enclosure:

GRP to IP65 protection

Analyser Dimensions:

620 (h) x 500 (w) x 280 (d) mm

Weight:

35 Kg

Controller Dimensions:

"Protoc® 300" has an integral controller
"Micro Spyder®" 174(h) x 200(w) x 224(d) mm
"Spyder®" 295(h) x 415(w) x 245(d) mm

Outputs:

Analogue 4-20mA standard (each channel)

 

Various process & utility alarms (model dependent)

Power supply:

110/230VAC 50/60Hz


Features
Fast response from 3 minutes
Two point auto-calibration
Optional automatic cleaning
Continuous multi-point analysis
Telemetry outputs
Optional data logging & Windows compatible software
top of the page

  Tel: 01732 882044      e-mail: toc@pollution-ppm.co.uk