PROTOC® TL is a fast, reliable and
flexible on-line analyser system designed to accurately determine
trace level TOC/TC concentrations in real time. The analyser comprises
a wall mount, analyser section typically supplied complete with
a controller providing local display and outputting signals.
- Protoc® 100
Low cost analyser with manually operated valve to calibrate
- Protoc® 300
Economical analyser for single stream analysis with auto-zero
- Protoc® microSpyder and Web(s)
Economical multi-point 1-4 stream analysis with touch screen
controller
- Protoc® Spyder and Web(s)
Advanced multi-point 1-4 stream analysis with touch screen controller
PROTOC® instruments use an Ultra-Violet
promoted persulphate oxidation to continuously determine the contamination
of organic chemicals present in the sample. A carrier gas continuously
sparges the reaction vessel liberating the resultant CO2 gas which
is delivered to a low level, Infra-Red detection system. Inorganic
carbonates can be automatically removed by sample pre-treatment
using an acid sparge.
|
 |
 |
Various models are available offering different levels of
automation. Automatic zero point calibration maintains accuracy
and a sample / calibration standard injection port enables simple
validation of measurement performance. Various options are also
available for data display, data logging and telemetry communications. |
Measuring Range: | 0-5 up to 50,000 PPM (with dilution) |
Repeatability: | +/- 2% or better |
Enclosure: | GRP to IP65 protection |
Analyser Dimensions: | 620 (h) x 500 (w) x 280 (d) mm |
Weight: | 35 Kg |
Controller Dimensions: | "Protoc® 300" 212(h) x 252(w) x 130(d) mm
"Micro Spyder®" 174(h) x 200(w) x 224(d) mm "Spyder®" 295(h) x 415(w) x 245(d) mm |
Outputs: | Analogue 4-20mA standard (each channel) |
| | Various process & utility alarms (model dependent) |
Power supply: | 110/230VAC 50/60Hz |
|